IESE G88 Sensor Responses

Coil and shunt resistance values affect the sensitivity, damping and pole values. To calculate poles, zeros and sensitivities for G88s built with other resistor values, click here.

Note:

IESE does not list current damping or shunt resistor information, so responses the G88 are given with a calculated shunt resistance that assumes a total damping of 0.7071 critical. Additional parameters needed to customize these responses using the link above are as follows:
Fundamental Frequency Mass Open Circuit Damping Intrinsic Sensitivity
4.5 Hz 0.023 Kg 0.27 78.74 V/M/S
2.0 Hz 0.023 Kg 0.61 78.7 V/M/S

These nominal responses are valid when the input resistance of the attached amplifier or datalogger is at least 2 orders of magnitude larger than the total resistance of the disconnected sensor. If this is not the case, total damping will not be 0.7071*critical and the poles will not describe the system accurately.

Manufacturer Model Frequency / Period Coil Resistance Shunt Resistance Sensitivity RESP Filename Source Comments
IESE G88 4.5 Hz / 0.22 s 3810 Ohms 7096 Ohms 51.23 V/M/S RESP.XX.NS375..SHZ.G88.3180.7096.4_5.51_23 IESE
IESE G88 2.0 Hz / 0.22 s 3810 Ohms 106661 Ohms 76.0 V/M/S RESP.XX.NS376..SHZ.G88.3180.106651.2.76 IESE

Mary Templeton 5/6/2016