Nominal Response Library | |
---|---|
Element | sensor |
Manufacturer | Geotech |
Model | S-13J |
These responses assume a damping factor of 0.707 and natural frequency of 1 Hz. For other damping factors or natural frequencies, the poles need to be recalculated as follows:
p1, p2 = (-2 * pi * f0 * bt) +/- j * (2 * pi * f0 * sqrt(1 - bt**2) )
where - p1, p2 = poles - f0 = natural frequency - bt = damping factor
Configurations
Updating...
Showing 1 of 1
Long-Period Corner
|
Sensitivity
|
Coil Resistance
|
Sensor Type
|
Formats |
---|---|---|---|---|
sensor_Geotech_S-13J_LP1_SG344_RC6400_STgroundVel | ||||
1 s | 344 V/m/s | 6400 | groundVel |